Mallinckrodt INOmax DSIR Plus MRI Operation Manual page 137

Table of Contents

Advertisement

Environmental
Temperature:
Humidity:
Ambient
Pressure:
Water Ingress
Protection:
INOMAX Regulator
Inlet Pressure:
Outlet Pressure:
Cylinder Valve Connector: CGA 626
Electrical
Important: Disconnect main power cord to isolate equipment from main power.
Input Voltage:
Input Power:
Input Fuse:
Classification:
Standards:
Battery Backup:
USB Port:
Ethernet Port:
RS 232:
Infrared Port:
Alarm Log
The alarm history is deleted when device is turned off. However, the service log, which is
accessible by service personnel is maintained (including alarm log) when power is cycled
and/ or when total power loss occurs.
Part No. 20568 Rev-02
2015-08
Operating:
Transport/Storage:
5 to 40°C
-20 to + 60°C
15 to 95% RH
15 to 95% RH
non-condensing
non-condensing
57 to 110 kPa
57 to 110 kPa
IPX1
14 to 155 Bar (200 to 2,248 psig)
1.7 to 2.4 Bar (25 to 35 psig)
100-240 V AC @ 50 / 60 Hz
110 VA max
3 A
Class I, Type B
CSA certified to meet the following for medical electrical equipment:
• UL 60601-1: 2003 edition 2
• ANSI/AAMI 60601-1: 2005 edition 3
• IEC 60601-1: 2005 edition 3
A sealed lithium ion rechargeable battery provides power backup
to operate the system for up to six hours when fully charged.
Connect the system to an electrical outlet for at least ten
hours to charge the battery.
When the low battery alarm occurs, there are 30 minutes
until battery depletion.
Dispose of used batteries according to local regulations.
Disabled.
For service only. Not for use when patient is connected.
Enables serial communications for use with hospital electronic health record
(EHR) system.
Infrared communication with the INOMAX cylinder.