Mallinckrodt INOmax DSIR Plus MRI Operation Manual page 149

Table of Contents

Advertisement

8c
8e
8b
8a
Part No. 20568 Rev-02
2015-08
7. Performance Test:
8d
8. INOblender Test:
8e
a. Verify O
flowmeter is set to 10 L/min.
2
b. Set INOMAX dose to 40 ppm.
c. Verify monitored values:
Acceptable NO Value (ppm)
Acceptable NO
Value (ppm)
2
Acceptable FiO
(%)
2
d. Set INOMAX dose to 0 ppm.
• "Set Dose is Zero, Please Close Cylinder
Valve" reminder will appear- DO NOT
close cylinder valve at this time, dismiss
reminder.
e. Turn oxygen flowmeter OFF.
a. Remove injector module from pre-use
assembly and reconnect tubing.
b. Remove O
tubing from flowmeter and attach
2
to INOblender outlet.
c. Set INOblender flow to 10 L/min, INOMAX
dose to 40 ppm.
d. Verify monitored values on the
INOmax DS
Plus MRI.
IR
Acceptable NO Value (ppm)
e. Set INOblender dose and flow to 0.
35 - 45
< 1.5
95 ± 3
32 - 48
8-3