2/ Automated Pre-Use Checkout
Connect the INOmax DS
be connected to an electrical outlet to maintain a full battery charge.
Keep the power cord off of the ground and away from moving parts.
Caution:
Part No. 20568 Rev-02
2015-08
Plus MRI power cord to a hospital-grade AC outlet. The power cord must always
IR
1. Turn ON INOmax DS
An INOmax DS
Plus MRI splash screen will
IR
appear once the device is turned ON followed by
the speaker sounding.
Low calibration automatically starts
Note:
following the INOmax DS
self test.
2. Press the NEXT button to review the warnings
(steps 1 to 3).
• The INOmax DS
WARNING:
classified as MR Conditional
with MR scanners of 1.5 or 3.0
Tesla strength ONLY in areas
where the field strength is less
than 100 gauss.
• Do not exceed 100 Gauss;
system operation may be
impacted.
• Confirm cart auto-brake
function. Optionally connect
tether.
3. Test the auto-brake.
Pull the yellow brake handle up, the
Note:
INOmax DS
IR
move freely. Release the handle, the cart
should not move when pushed. If the cart
fails to move or moves when the brake
handle is not pulled, contact your local
representative.
• If the cart fails to move when
WARNING:
the brake handle is pulled,
or moves when the brake
handle is not pulled, do not
use the INOmax DS
and contact your local
representative.
Plus MRI.
IR
Plus MRI
IR
Plus MRI is
IR
Plus MRI cart should now
Plus MRI
IR
2-1