Emptying the Water Bottle
The water bottle (see Figure 3-8) collects fluids separated from the patient gas sample.
• Empty and clean the water bottle before each patient use and empty whenever the bottle is more than half
full.
• Empty the water bottle routinely. Allowing it to fill and overflow may cause system errors.
• A "Water Bottle Full" message will remind the user to empty and clean the bottle should it become full.
Note:
Monitoring will be temporarily interrupted when the "Water Bottle Full" message is indicated.
Figure 3-8
During delivery of INOMAX to a patient
Note:
1. The disposable water separator cartridge on the rear of the water bottle housing protects the
monitoring system from moisture and other contaminants and may need to be replaced
occasionally while in use (refer to Section 6/ Maintenance).
2. To avoid medications interfering with the gas monitoring system, administer any aerosolized
medications distal to the sampling tee in the breathing circuit.
Part No. 20568 Rev-02
2015-08
To empty the Water Bottle:
1. Remove the bottle by pulling it straight down
(see Figure 3-8).
2. Discard the contents according to an
approved fluid waste disposal policy.
3. Clean the bottle.
4. Replace the bottle by pushing it up into
position.
5. Check for leaks by running the system and
occluding the sample line until the "Sample
Line/Filter Block" alarm message appears.
3-15