Mallinckrodt INOmax DSIR Plus MRI Operation Manual page 136

Table of Contents

Advertisement

NO Delivery
Set NO Range:
Set NO Resolution:
Accuracy @ 20°C:
NO Inlet Pressure:
Maximum NO Supply Pressure:
NO Low Pressure Alarm:
Max Circuit Pressure:
Breathing Circuit Gas Composition:
MR Injector Module
Conical Connectors:
Autoclavability:
Maximum Pressure Drop:
Gas Monitoring
Gas
Nitric Oxide:
Nitrogen Dioxide:
Oxygen:
Max Breathing Circuit Pressure
Calibration
Rise Time
Sample Flow
Integrated Pneumatic Backup Delivery
Integrated Pneumatic Backup Delivery = 250 mL/min Fixed Flow of NO/N
Physical
Delivery System (including: INOmax DSIR Plus MRI, INOblender,
Nominal Weight:
Nominal Width and Depth:
Nominal Height:
7-2
Range
0 - 10 ppm
10 - 100 ppm
0 - 10 ppm
18 - 100 % v/v
:
:
:
:
cart, 2 regulators and 2 INOMAX cylinders)
80 kg (176 lb)
53 cm (21 inch) W x 56 cm (22 inch) D
149 cm (58 inch) H
0.1 - 80 ppm ( 800 ppm cylinder )
0.1 ppm from 0 to 1 ppm
1 ppm from 1 to 40 ppm
2 ppm from 40 to 80 ppm
± 20% or 2 ppm, whichever is the greater
1.7 to 2.4 Bar (25 to 35 psig)
2.4 Bar (35 psig)
1.6 Bar (23 psig) (nominal)
1.4 Bar (20 psig)
Air / O
mixtures
2
Inlet, 22 mm female.
Outlet, 22 mm male and 15 mm female.
Autoclavable at 134°C for 3 minutes at 27 psig.
1.5 cmH
O at 60 L/min
2
Resolution
Accuracy
0.1
± (20% of reading + 0.5 ppm)
1
± (10% of reading + 0.5 ppm)
0.1
± (20% of reading or 0.5 ppm
whichever is greater)
1
± 3% v/v
150 cmH
0
2
Daily zero; span when needed
30 seconds (10 - 90 %)
230 mL/min
2
Part No. 20568 Rev-02
2015-08

Advertisement