Part No. 20568 Rev-02
2015-08
3. When the low calibration is successful, a single
tone will be heard and the main screen will
appear. A two minute Monitor Alarm Delay
Active indicator will occur, preventing monitoring
alarms from occurring while the measured value
stabilizes. All system alarms are still active.
If the low calibration was unsuccessful, the
INOmax DS
Plus MRI will automatically
IR
attempt another low calibration. If the second low
calibration attempt fails, the alarm area will display
a Low Calibration Failed alarm.
• Attempt a manual low calibration.
If a sensor has failed, the display will indicate
the failed sensor symbol in the monitoring
area of that sensor. (Press the
Alarm Help button for on-screen
alarm help or see Section 4/ Alarms
and Troubleshooting.)
5-3