Figure 3-7 Correct Position
Incorrect Position
Part No. 20568 Rev-02
2015-08
3. Insert the NO/N
quick-connect fitting into
2
the purge port on the back of the
INOmax DS
Plus MRI and firmly push until
IR
the regulator pressure gauge reads zero (this
purges any NO
that has accumulated in the
2
hose and regulator).
All INOmax DS
WARNING:
devices must be purged
before use to ensure the
patient does not receive
excess of NO
4. Prior to connecting a regulator hose,
ensure the inlet connectors, on the
INOmax DS
Plus MRI unit, have the
IR
knurled sleeve set in the back position
(toward the INOmax DS
see Figure 3-7).
Plus MRI
IR
.
2
Plus MRI unit,
IR
3-11