All of these actions can be performed while delivering INOMAX to the patient:
Alarm
9. Delivery Stopped
10. Drug Past Expiry Date
WARNING: Delivery
Stopped will occur two
minutes from point when
Drug Past Expiry Date
alarm is activated.
Drug Concentration
11.
Mismatch
WARNING: Delivery
Stopped will occur two
minutes from point when
Cylinder Concentration
Mismatch alarm is
activated.
Part No. 20568 Rev-02
2015-08
High Priority Alarms
Possible Cause
MONITORED NO > 100 ppm
for at least 12 seconds.
Drug past expiry date.
Drug concentration mismatch.
INOmeter may have failed.
INOMAX cylinder valve is
closed.
INOMAX cylinder is not
detected.
INOMAX cylinder is expired.
INOMAX cylinder is the wrong
concentration.
Recommended Action
Manually ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON.
Replace INOMAX cylinder.
Replace INOMAX cylinder.
Replace INOMAX cylinder.
Open INOMAX cylinder valve.
1. Replace INOMAX cylinder.
2. Contact Technical Support.
1. Close expired cylinder valve.
2. Remove expired INOMAX cylinder from
INOmax DS
Plus MRI cart.
IR
3. Connect an INOMAX cylinder with a valid
expiration date.
4. Contact Technical Support.
1. Close mismatched cylinder valve.
2. Remove wrong concentration INOMAX
cylinder from INOmax DS
3. Connect an INOMAX cylinder with a valid
concentration to INOmax DS
4. Contact Technical Support.
Plus MRI cart.
IR
Plus MRI.
IR
4-7