Mallinckrodt INOmax DSIR Plus MRI Operation Manual page 112

Table of Contents

Advertisement

5-10
When using the high calibration wizard, if the
BACK button is pressed during the NO high
calibration process, the progress bar will turn
yellow, signifying a cancelled calibration.
• Press the NEXT button to start the high
calibration wizard.
• Press the START CAL button to start the
calibration again.
• Press the CANCEL button to exit the NO high
calibration screen.
If the calibration was unsuccessful, the NO
progress bar will turn red.
• Attempt another calibration.
To repeat the NO high calibration, press
Note:
the START CAL button at the bottom of
the screen.
If the NO sensor has failed, the display indicates
the failed sensor symbol in the monitoring area
of that sensor.
Repeat calibration. (Press the Alarm
Help button for on-screen alarm
help or see Section 4/ Alarms and
Troubleshooting.)
Part No. 20568 Rev-02
2015-08