Pre-Use Assembly
1
10 LPM
Off
2
1. O
Flowmeter
2
2. O
Tubing
2
3. 15M x 4.5 mm Adapter
4. 22M / 15F x 22M / 15F Adapter
5. Injector Module
6. 300 mm of 22 mm hose
7. Patient Gas Sample Line with Nafion
8. Gas Sample Tee
9. Injector Module Electrical Cable
10. NO/N
Injector Tube
2
Figure 8-1

Additional Dose Setting Information

Each click on the control knob corresponds to
a known change in dose. The incremental dose
per click corresponds to a value dependent upon
the dose range in which the change is made, as
illustrated in the table at right.
8-4
3
5
4
10
9
Dose Setting Range Dose Change Per Click
7
6
1 to 40 ppm
40 to 80 ppm
8
1 ppm
2 ppm
Part No. 20568 Rev-02
2015-08