5. Select the dose button on the screen. On the
INOmax DS
Plus MRI, rotate the control wheel
IR
to set the NO dose.
6. Confirm the change by pressing the Control
wheel or dose button on the screen.
7. Set the user-adjustable alarm settings on the
INOmax DS
Plus MRI and on the ventilator or
IR
breathing system.
The first time a dose is set from zero,
Note:
the upper and lower NO alarm limits
are set 50% above and 50% below
the set dose.
Part No. 20568 Rev-02
2015-08
3-5