Manual Pre-Use Checkout - Mallinckrodt INOmax DSIR Plus MRI Operation Manual

Table of Contents

Advertisement

8/ Appendix

Manual Pre-Use Checkout

• Only use a size "88" (1,963 liters) cylinder that is marked "MR Conditional.
WARNING:
Keep cylinder at 100 gauss or less." with the INOmax DS
scanner room. Use of any other cylinder may create a projectile hazard.
• The INOmax DS
1.5 or 3.0 Tesla strength ONLY in areas where the field strength is less than 100
gauss.
The following instructions are provided for when the on-screen pre-use wizard is not used.
2h
2f
2e
2g
Part No. 20568 Rev-02
2015-08
Plus MRI is classified as MR Conditional with MR scanners of
IR
1. Turn device ON, low calibration will begin and
complete (Continue with steps 2-4 while calibration
completes).
2. Initial Connections:
Confirm attachment of the following:
a. Water separator cartridge, water bottle, and MR
conditional patient gas sample line in place.
b. MR conditional injector module cable and tubing
are connected.
c. Plug in power cord and verify AC power
2a
light is ON.
d. Regulator to INOMAX cylinder.
e. Regulator hose to INOmax DS
f. INOblender hose connected and white lock in
place.
g. Oxygen source (50 psig) to back of INOblender.
h. IR cable in place.
3. Assemble pre-use set-up connectors (see Figure
8-1). Do Not turn on O
4. High Pressure Leak Test:
Open/close INOMAX cylinder valve.
a. Verify, at least 34.5 bar (500 psig) cylinder
pressure.
b. Verify, no decrease in cylinder pressure for
30 seconds.
Plus MRI while in the
IR
Plus MRI inlet.
IR
flowmeter yet.
2
8-1