Replacing the O
Figure 6-6
Figure 6-7
NO
Shorting Wire
2
6-8
, NO and NO
2
Sensors
(cont'd)
2
Replace the sensor cover and tighten the two
screws clockwise (see Figure 6-6).
Note:
Newly Installed Sensor
O
and NO
2
2
NO
Insufficient conditioning will result in inaccurate
gas readings.
Perform a low and high calibration for the sensor
before returning the system to use.
Time to Condition Prior
to Calibration
40 minutes
5 hours
Part No. 20568 Rev-02
2015-08