5. Open the cylinder valve on the new cylinder
(this may activate the "Two Cylinders Open"
alarm until the empty cylinder valve is closed).
6. Close the cylinder valve on the empty
cylinder and remove the supply line from
the back of the INOmax DS
7. Depressurize by using the purge port on
the back of the INOmax DS
to removing the regulator from the empty
cylinder.
8. Replace the empty cylinder with a full cylinder
on the cart.
Plus MRI.
IR
Plus MRI prior
IR
Part No. 20568 Rev-02
2015-08