Mallinckrodt INOmax DSIR Plus MRI Operation Manual page 135

Table of Contents

Advertisement

7/ Product Specifications
• The target patient population is controlled by the drug labeling for INOMAX and
WARNING:
is currently neonates. The INOmax DS
other patient populations.
• Outside of the United States, use of the INOmax DS
use in accordance with INOMAX nitric oxide for inhalation prescribing information as
established with the national health authority.
The INOmax DS
Plus MRI is designed to function in the parameter ranges listed in this section.
IR
Use outside of these ranges is not recommended.
Ventilator Compatibility
The INOmax DS
Plus MRI is compatible with the ventilators listed below:
IR
Ventilators/Breathing Systems validated for use in the United States
Manufacturer
Bio-Med Devices
CareFusion (formerly Pulmonetic
Systems)
Nasal Cannula
Maquet
Part No. 20568 Rev-02
2015-08
Inspiratory Flow Rate:
Respiratory Rate:
Airway Peak Pressure:
PEEP:
Model
MVP-10
LTV 1200 MRI
Salter Labs, Infant, REF 1601,
with 7 foot supply tube (or
equivalent)
Servo-i
Plus MRI is not intended to be used in
IR
Plus MRI is limited to the
IR
Measure
L/min
bpm
cmH
O
2
cmH
O
2
MR Conditional
Specification
2 - 120
6 - 60
0 - 70
0 - 20
7-1