Mallinckrodt INOmax DSIR Plus MRI Operation Manual page 115

Table of Contents

Advertisement

Part No. 20568 Rev-02
2015-08
After reaching step 7 of the high calibration
wizard, the calibration will take approximately
three minutes, during which a progress bar for the
sensor indicates progress.
If the CANCEL button is pressed during
Note:
the NO
high calibration process, the
2
calibration will be cancelled and the user
will be returned to the high calibration
screen.
When the NO
high calibration is successful, a
2
single tone will be heard and the progress bar will
turn green. Press the NEXT button to return to the
high calibration screen.
The monitor displays should indicate
Note:
approximately 21% O
0.0 ppm NO.
10. Disconnect the patient gas sample line from
the calibration tubing and close the INOcal
cylinder valve.
A Monitor Alarm Delay Active indicator will occur
anytime the user exits the calibration screen,
even if a calibration was not initiated. All other
alarms are still active (see Section 4/ Alarms and
Troubleshooting for additional information).
, 10 ppm NO
and
2
2
5-13