Cleaning the INOmax DS
• Do not autoclave or gas sterilize the INOmax DS
Caution:
• Do not clean with the power connected and the INOmax DS
• Be sure that the INOmax DS
• Do not saturate the INOmax DS
the system and damage internal components.
• Do not use organic, petroleum based solvents, glass cleaners, acetone or other harsh
cleaning agents.
• Do not use abrasive cleaning agents (such as steel wool, silver polish or cleaner).
• Do not touch or rub the display panel with abrasive cleaning compounds or anything which
can scratch the panel.
• Do not use organic solvents to clean the display panel.
Cleaning Procedure
Apply cleaning agent to a cloth before application; do not spray directly on the delivery
Caution:
system to prevent pooling and direct contact with electrical connections which can cause
damage over time.
External surfaces and the Display panel
• Disconnect the power cord from the wall outlet and turn the INOmax DS
• Clean the outer surface of the INOmax DS
water solution, isopropyl alcohol (70%) or with one of the following cleaning agents while following the
manufacturer's recommendations.
Cleaning Agent
Precise Hospital Foam Cleaner Disinfectant by
Caltech Industries
Pure Green 24 by Pure Green, LLC
Part No. 20568 Rev-02
2015-08
Plus MRI
IR
Plus MRI.
IR
Plus MRI is completely dry before using.
IR
Plus MRI with excessive solution. Liquid may flow into
IR
Plus MRI with a soft cloth dampened in a mild soap and
IR
Active Ingredients
o-Phenylphenol < 0.37%
Other ingredients 99.63%
SDC – silver ions 0.003%
Citric acid 4.84%
Other ingredients 95.157%
Plus MRI turned ON.
IR
Plus MRI OFF before cleaning.
IR
6-3