INOmeter Operation

• The INOmeter is a time-metric device which records the amount of time the INOMAX cylinder valve
is opened.
• When used with INOmax DS
the INOmax DS
Plus MRI and the INOmeter. The INOmeter communicates the INOMAX cylinder
IR
concentration and the expiration date to the INOmax DS
and dose information are communicated from the INOmax DS
Figure 1-13
Figure 1-14
1-22
Plus MRI, two-way infrared (IR) communication occurs between
IR
Plus MRI. Patient ID (when confirmed)
IR
Plus MRI to the INOmeter.
IR
• Cylinders are shipped with the
Note:
INOmeter covered in a tamper-proof
seal.
• A valve lock is secured to the cylinder
by a lanyard.
• The lock must be removed to open
the cylinder valve for use.
1. Remove and properly dispose of tamper-proof
seal or covering (see Figure 1-13).
2. The lock is secured to the cylinder by a
lanyard (see Figure 1-14).
Part No. 20568 Rev-02
2015-08