Mallinckrodt INOmax DSIR Plus MRI Operation Manual page 84

Table of Contents

Advertisement

Alarm Help
All of these actions can be performed while delivering INOMAX to the patient:
Alarm
1. High NO
4-2
High Priority Alarms
Possible Cause
The High NO alarm level may
be inappropriately set.
Circuit setup incorrect.
NO sensor may require
calibration.
MR injector module may not
be functioning properly.
NO sensor may require
calibration.
Recommended Action
Confirm High NO alarm limit is set appropriately.
Check circuit for correct setup.
Perform low calibration.
1. Manually ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON
2. Replace the MR injector module (INOMAX
delivery will be interrupted).
3. Replace the MR injector module cable.
(INOMAX delivery will be interrupted).
Perform a high calibration outside of the MR
scanner room.
1. Perform high NO calibration.
2. Contact Technical Support.
Part No. 20568 Rev-02
2015-08