Infrared Communication between the INOMAX Cylinders and
the INOmax DS
Plus MRI
IR
Loss of communication between the INOmax DS
WARNING:
for more than one hour will result in interruption of INOMAX delivery.
The INOmax DS
Plus MRI has an interface using infrared (IR) technology which allows the
IR
INOmax DS
Plus MRI to communicate with the INOmeter (which is mounted to each INOMAX cylinder).
IR
The INOmax DS
Plus MRI checks the INOMAX cylinder for the correct expiration date and cylinder
IR
concentration. The INOmax DS
any open INOMAX cylinder.
The INOmax DS
Plus MRI cart (PN 10076) has a cover (see Figure 1-12,
IR
mounted directly above each INOMAX cylinder. When INOMAX cylinders are loaded, communication will
take place between the INOmax DS
phase of the INOmax DS
an INOMAX cylinder is recognized by the INOmax DS
INOmax DS
Plus MRI Cart", page 1-21).
IR
Caution:
Nothing should be placed between the INOmeter and the cart to which it is attached.
Magnetic Interference
A strong magnetic field such as that from an MRI system can affect the ability of
WARNING:
the INOmeter to detect if the cylinder valve is open. This can cause a "Cylinder
Valve Closed" alarm to occur when the cylinder valve is actually open. If this alarm
occurs, reposition/rotate the INOmax DS
to reduce the magnetic interference in the area of the INOmeter until the cylinder
handle graphic on the display turns green. This will resolve the "Cylinder Valve
Closed" alarm. Typically the required INOmax DS
is less than 6 inches (15 cm) / 90 degrees. Note that Interruption of INOMAX therapy
will occur one hour from point when the "Cylinder Valve Closed" alarm is activated if
the alarm is not resolved.
Communication Interference
The INOmax DS
Plus MRI transceiver is located under the cart cover and should be protected from outside
IR
interference. The INOmax DS
IR energy sources. The INOmax DS
projecting towards the floor (see dotted lines in Figure 1-12). The specifications of the IR beam call for it to
have a range of 20 cm (7.9 in). Based on these specifications it should not affect other devices in the vicinity
of the INOmax DS
Plus MRI.
IR
The INOmeter uses a lower energy source which results in a lower IR beam range than the
INOmax DS
Plus MRI cart. The INOmeter does not transmit IR signals unless it is mounted on the
IR
INOmax DS
Plus MRI cart.
IR
Part No. 20568 Rev-02
2015-08
Plus MRI also transmits a confirmed patient identifier to the INOmeter on
IR
Plus MRI and the INOmeter (see Figure 1-12,
IR
Plus MRI is complete. A cylinder icon will be displayed on the main screen when
IR
Plus MRI cart was designed to protect the INOmeter from external light/
IR
Plus MRI transceiver transmits via a 30 degree transmission cone
IR
Plus MRI and the INOmeter
IR
1
) with an infrared transceiver
Plus MRI (see "Loading INOMAX Cylinders onto the
IR
Plus MRI cart outside the 100 Gauss area
IR
Plus MRI cart location adjustment
IR
2
) after the boot up
1-19