Part No. 20568 Rev-02
2015-08
After reaching step 7 of the high calibration
wizard, the calibration will take approximately
three minutes, during which a progress bar for the
sensor indicates progress.
If the CANCEL button is pressed during
Note:
the NO high calibration process, the
calibration will be cancelled and the user
will be returned to the high calibration
screen.
When the NO high calibration is successful, a single tone
will be heard and the progress bar will turn green. Press
the NEXT button to return to the high calibration screen.
The monitor displays should indicate
Note:
approximately 0.0% O
45 ppm NO.
10. Disconnect the patient gas sample line from
the calibration tubing and close the INOcal
cylinder valve.
A Monitor Alarm Delay Active indicator will occur anytime
the user exits the calibration screen, even if a calibration
was not initiated. All other alarms are still active (see
Section 4/ Alarms and Troubleshooting for additional
information).
, 0.0 ppm NO
and
2
2
5-9