Using The Inomax Ds Ir Plus Mri In The Mr Scanner Room; Transferring To And From The Mr Scanner Room - Mallinckrodt INOmax DSIR Plus MRI Operation Manual

Table of Contents

Advertisement

Using the INOmax DS
• If the cart fails to move when the brake handle is pulled or moves when the brake
WARNING:
handle is not pulled, do not use the INOmax DS
representative.
• The gauss alarm will sound if the INOmax DS
the MR scanner. If alarm sounds, move system away from the MR scanner until
the gauss alarm stops sounding.
• Do not perform a high calibration procedure in MR scanner room, calibration
equipment is a potential projectile hazard.

Transferring to and from the MR scanner room

Prior to moving the patient into the MR scanner room, complete the following steps:
1. Complete a pre-use procedure on the INOmax DS
properly.
2. Match the dose and alarm settings on the INOmax DS
transported the patient to the MR scanner room.
3. Remove the INOmax DS
the INOmax DS
Plus MRI injector module. Repeat the same steps to change the patient gas sample
IR
line (for more detailed information, see page 3-4).
4. Move the patient, the MR ventilator and INOmax DS
5. Position the INOmax DS
The gauss alarm will sound if the INOmax DS
WARNING:
MR scanner. If alarm sounds, move system away from the MR scanner until the
gauss alarm stops sounding.
6. The tether attachment point allows for a facility supplied cable to be secured to the cart as a redundant
means to limit the distance the cart can move.
7. Keep the INOmax DS
8. Move the patient, the MR ventilator and INOmax DS
9. Remove the INOmax DS
insert the INOmax DS
line.
10. Verify ventilator and INOmax DS
3-2
Plus MRI in the MR Scanner Room
IR
Plus injector module from the patient ventilator circuit and immediately insert
IR
Plus MRI outside of the MR exclusion zone (see Figure 3-1).
IR
Plus MRI outside of the MR exclusion zone.
IR
Plus MRI injector module from the patient ventilator circuit and immediately
IR
Plus injector module. Repeat the same steps to change the patient gas sample
IR
Plus function following transition from the MR scanner room.
IR
Plus MRI and contact your local
IR
Plus MRI system is too close to
IR
Plus MRI to verify the system is functioning
IR
Plus MRI with the INOmax DS
IR
Plus MRI into the MR scanner room.
IR
Plus MRI system is too close to the
IR
Plus MRI outside of the MR scanner room.
IR
Plus that has
IR
Part No. 20568 Rev-02
2015-08