Mallinckrodt INOmax DSIR Plus MRI Operation Manual page 110

Table of Contents

Advertisement

6a
6
9c
9b
5-8
7
To INOmax DS
Plus MRI
IR
9a
6. Open the INOcal cylinder valve (turn counter-
clockwise) to flow gas.
6a. If the pressure is in the red or black zone (0-
25 psig) select another INOcal cylinder.
7. Attach tubing kit to patient gas sample line.
If the date is flashing on the high
Note:
calibration screen, the calibration is
past due.
8. From the high calibration screen
(third menu level), press the 45
ppm NO button to initiate the NO
high calibration.
9a. To continue the high calibration wizard, press
the NEXT button.
9b. To start the high calibration without using the
wizard, press the START CAL button.
9c. To exit the NO high calibration, press the
CANCEL button.
Part No. 20568 Rev-02
2015-08