5-14
When using the high calibration wizard, if the
BACK button is pressed during the NO
calibration process, the progress bar will turn
yellow, signifying a cancelled calibration.
• Press the NEXT button to start the high
calibration wizard.
• Press the START CAL button to start the
calibration again.
• Press the CANCEL button to exit the NO
calibration screen.
If the calibration was unsuccessful, the NO
progress bar will turn red.
• Attempt another calibration.
To repeat the NO
Note:
the START CAL button at the bottom of
the screen.
If the NO
sensor has failed the display indicates
2
the failed sensor symbol in the monitoring area
of that sensor.
Repeat calibration. (Press the alarm
help button for on-screen alarm
help or see Section 4/ Alarms and
Troubleshooting.)
high
2
high
2
2
high calibration, press
2
Part No. 20568 Rev-02
2015-08