Testing the GaussAlert Function

• Keep the test magnet tool away from pacemakers, ICDs and other implanted
WARNING:
medical devices.
• If either GaussAlert fails testing (does not alarm when the magnet tool is used),
contact the Authorized Representative listed on the back cover of the operation
manual to request a replacement INOmax DS
• Do not use the INOmax DS
• Do not use or store the test magnet tool in the MR scanner room.
Caution:
• Keep the test magnet tool away from the INOmax DS
magnets can damage computer monitors, watches, pulse oximeters, and other mobile
handheld devices.
• Keep the test magnet tool away from magnetic media such as credit cards, magnetic I.D.
cards, cassette tapes, video tapes or other such devices.
Figure 6-1
6-2
Plus MRI if neither gauss alarm is functional.
IR
1. Use the GaussAlert test magnet tool
(PN 50192) to test alarm function.
2. Place the test magnet tool on the GaussAlert
alarm in the area identified in Figure 6-1.
3. The alarm will activate and a short flash (less
than two seconds) of the LED indicates proper
function.
4. Move the test magnet tool away from the
GaussAlert until the alarm stops sounding.
5. Repeat this procedure on the second
GaussAlert.
Note:
Magnet Tool
Plus MRI system.
IR
Plus MRI user screen. Neodymium
IR
If the LED stays illuminated the entire
time the gauss alarm is ON or does
not illuminate at all, the battery needs
to be replaced. Contact the Authorized
Representative listed on the back cover
of the operation manual to request a
replacement INOmax DS
system.
Part No. 20568 Rev-02
Plus MRI
IR
2015-08