All of these actions can be performed while delivering INOMAX to the patient:
Alarm
3. High NO
2
4. High O
2
4-4
High Priority Alarms
Possible Cause
The High NO
alarm setting
2
may be inappropriately set.
The patient circuit setup may
be incorrect.
-Patient circuit flow
interrupted.
Incomplete system
purge
Monitored NO
value is too
2
high.
NO
sensor may require
2
calibration.
The High O
alarm setting
2
may be inappropriately set.
O
sensor may require
2
calibration.
Recommended Action
Confirm High NO
alarm limit is set
2
appropriately.
Check circuit for correct setup.
Repeat the purge procedure with the MR
injector module out of the patient breathing
circuit.
Use INOblender if necessary.
Consider reducing the INOMAX dose according
to INOMAX prescribing information.
1. Perform low calibration.
2. Perform a high NO
calibration outside of the
2
MR scanner room.
3. Contact Technical Support.
Confirm High O
alarm limit is set appropriately.
2
1. Perform low calibration.
2. Perform a high O
calibration outside of the
2
MR scanner room.
3. Contact Technical Support.
Part No. 20568 Rev-02