Initialization mode selection

The following initialization (init) modes are selectable:
Table 22
Initialization Mode Switches
1
Requires firmware B.06.40 or above. Modules without LAN on board, see G1369C LAN Interface Card
Bootp
When the initialization mode Bootp is selected, the module tries to download
the parameters from a Bootp Server. The parameters obtained become the
active parameters immediately. They are not stored to the non-volatile
memory of the module. Therefore, the parameters are lost with the next power
cycle of the module.
Figure 58
Agilent 1200 Infinity Series DAD User Manual
SW 6
SW 7
OFF
OFF
OFF
OFF
OFF
ON
OFF
ON
ON
OFF
Bootp (Principle)

Initialization mode selection

SW 8
Init Mode
OFF
Bootp
ON
Bootp & Store
OFF
Using Stored
ON
Using Default
OFF
1
DHCP
13
LAN Configuration
227