13
LAN Configuration

Initialization Mode Selection

Initialization Mode Selection

The following initialization (init) modes are selectable:
Table 32
Default IP address for LAN is 192.168.254.11.
DHCP address is the module's LAN MAC address.
Using Stored
When initialization mode Using Stored is selected, the parameters are taken from
the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured previously
by one of the described methods.
Figure 59
Agilent InfinityLab LC Series Diode Array Detectors User Manual
Initialization Mode Switches
SW1
0
0
0
Note: The setting '0' (down) is essential.
Non-Volatile
RAM
Stored
Parameter
Using Stored (Principle)
SW2
SW3
SW4
SW5
0
0
0
0
0
0
0
1
0
0
1
0
Active
Parameter
SW6
Init Mode
0
Use Default IP Address
0
Use Stored IP Address
0
Use DHCP
209