Initialization mode selection

The following initialization (init) modes are selectable:
Table 16
Initialization Mode Switches
Bootp
When the initialization mode Bootp is selected, the module tries to download
the parameters from a Bootp Server. The parameters obtained become the
active parameters immediately. They are not stored to the non-volatile
memory of the module. Therefore, the parameters are lost with the next power
cycle of the module.
Figure 15
Agilent 1260 Infinity DAD and MWD User Manual
SW 6
SW 7
OFF
OFF
OFF
OFF
OFF
ON
OFF
ON
Bootp (Principle)

Initialization mode selection

SW 8
Init Mode
OFF
Bootp
ON
Bootp & Store
OFF
Using Stored
ON
Using Default
LAN Configuration