Verify Schedule

Use this function to schedule automatic analyzer verification.
Config
Verify Schedule
Schedule
Type
Day of
Week
Day of
Month
Hour
Duration
NOTE
Schedule Type
Select one choice from the list:
Never
Daily
Weekly
Monthly
Never
Daily
Weekly
Day of the Week
Monthly
Select the day of the week from the list.
Sunday
Monday
Tuesday
Day of the Month
Wednesday
Enter the 2-digit value for the day of the month.
Thursday
Friday
Saturday
Hour
Enter Value
Enter the 2-digit value for the hour of the day using
Enter Value
a 24-hour clock. Zero is the starting hour.
Duration
Enter Value
Enter the verify period value in minutes.
AMETEK recommends performing a verification cycle every week for
the first month of operation. After this start-up period, the analyzer
will have dried down sufficiently to increase the verification interval
to 1 month. AMETEK recommends that the duration of the verifica-
tion cycle be set to 60 (i.e. - 60 minutes).
|