Example 3: Setting the Low Alarm Limit

Task
Set the analyzer to produce a "low
concentration" alarm, when the moisture
concentration falls below 0 PPM.

Example 4: Enabling "Hold During Verify"

Task
Enable hold the analog outputs at the last
measured value, when the analyzer is Offline
to perform a Verification.

Example 5: Setting the High-End of the Analog Output

Task
Set up the analog output so that a moisture
concentration of 100 PPM produces a 20
current output.

Example 6: Setting the Low-End of the Analog Output

Task
Set up the analog output so that a moisture
concentration of 1 PPM produces a 4
current output.

Example 7: Switching to "Sensor Saver" Mode

Task
Set the analyzer to operate in the Sensor
Saver mode.
6-12 | 3050-OLV Moisture Analyzer
Action
Write one holding register (function 06).
Register address = 67 (holding register
#40068). Value = 0.
Action
Write one holding register (function 06).
Register address = 98 (holding register
#40099). Value = 1 (0-track during verify,
1-hold during verify).
Action
Write one holding register (function 06).
mA
Register address = 45 (holding register
#40046). Value = 100.
Action
Write one holding register (function 06).
mA
Register address = 46 (holding register
#40047). Value = 100 (multiplied by scale of
100).
Action
Write one holding register (function 06).
Register address = 99 (holding register
#40100). Value = 1 (enable sensor saver =1,
disable sensor saver = 0).
PN 305200901, Rev YT