Figure - Siemens SC 7000 Service Manual

Patient monitors
Table of Contents

Advertisement

SC 7000 and SC 9000XL Patient Monitors
17Leakage Current
Tests
ASK-T898-03-7600
7k9kXLSM.c4.CD_ROM.fm/04-99/kaupp
LEAKAGE
TESTER
LEAKAGE
TESTER
LEAKAGE
TESTER
.
F igure 4-4 Leakage Current Test Setups
Table 4-3 Leakage Current Tests
TEST
Combined Lead Leakage
Individual Lead Leakage
Paired Lead Leakage
Leakage with Line Voltage on Leads
SC 7000 and SC 9000XL monitors are battery operated devices,
functionally grounded through CPS or IDS power supply when operated
from a docking station. Leakage current tests assure that under both
normal and fault conditions, any leakage current does not exceed values
given
inTable
4-3.
1) Perform leakage current tests with monitor on a docking station, and
CPS, IDS Power Supply, or PSL plugged into leakage tester. See

Figure

4-4.
2) Follow leakage tester manufacturer's instructions to measure each of
leakage currents given in
• open ground
• reversed polarity
• open ground reversed polarity
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
!$
DOCKING
CPS
STATION
,'6
,'6
,
1),1,7<
'RFNLQJ
3RZHU
3RZHU
DOCKING
CPS
STATION
6XSSO\
6XSSO\
6WDWLRQ
,'6
3RZHU
36/
DOCKING
25
CPS
6XSSO\
STATION
Max. Current
<10µA
<10µA
<10µA
<50µA
Table
4-3, for each of following conditions:
Service Manual
MONITOR
MONITOR
MONITOR
73

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sc 9000xl

Table of Contents