Service Manual
5.1.2 Installing Ejection Shaft
Cover
5.1.3 Removing Right Side Panel
5.1.4 Installing Right Side Panel
5.1.5 Removing Left Side Panel
46
Caution
The cover has nubs (see
in the top and bottom sides of the shaft. Grasping the cover by
the top and bottom sides tends to increase the effectiveness of
1
the nubs. The increased force required to remove it can pull the
shaft completely out of the ejector mechanism, requiring depot
repair of the monitor.
3) Push ejector shaft back in, and remove PCMCIA card.
Use the procedure of
PCMCIA card ejector shaft cover.
Note: The button cover is not centered on the shaft, and can be
installed in only one orientation.
Figure 3-5Removing Left and Right Side Panels (SC 7000 left side panel
shown in illustration)
1) Place monitor left side down on a smooth clean surface.
2) With PCMCIA card ejector shaft cover removed, press firmly on dots
below each side of sync output connector (see narrow arrows in
3-5), to release latch.
3) While pressing on dots, slide panel downward (see broad arrows in
Figure
3-5) to release locking tabs, and then lift cover off of monitor.
1) With monitor left side down on a smooth clean surface, position side
panel locking tabs into slots in monitor housing, and press inwards and
upwards to lock side panel to monitor.
2) Reinstall PCMCIA card ejector shaft. See
1) Place monitor right side down on a smooth clean surface.
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors

in illustration at left) that grip grooves
Section
5.1.1, reversing the action in step 2, to install
/HIW
7k9kXLSM.c3.CD_ROM.fm/04-99/kaupp
5LJKW
Figure
Section
5.1.2.
ASK-T898-03-7600