Service Manual
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
80281xxx
HEMOPOD / HEMOMED - related
80282xxx
80283xxx
80284xxx
80289xxx
8028axxx
80291xxx
80292xxx
8029fxxx
Power -related
802a0xxx
802a1xxx
802a2xxx
802a3xxx
QRS Sync pulse - related
802a4xxx
802a5001
ERROR: Address: A Written:B Read:D
Mask:D (E F)
126
Description
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Cause
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Pod communications
Try disconnecting/
problem.
reconnecting cable.
Pod cable, Pod, or
If problem persists, swap
CPU board failed
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Replace Battery
Replace Main Processor
Board Subassembly.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
RAM test failed over
Isolate problem to CPS/IDS
TAXI in either CPS/
or Monitor. If monitor then
IDS or Monitor ;
replace Main Processor
Address of failure
Board Subassembly; If CPS
given in msg
or IDS, replace CPS or IDS.
Action
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp