CPS/IDS if monitor in CPS/IDS Mode.
8. Select alternative a, b, or c below to set Network Control appropriately
for customer site.
a If SC 7000 used only in Standalone mode, set Network Control to
OFF and power-cycle monitor. Verify that MAIN screen displays,
after completion of power-up sequence and proper selection for
New Patient Prompt. This completes Monitor setup.
b If SC 7000 is to be connected to an I
docking station --
b.1 Secure monitor on docking station.
b.2 Assure CPS/IDS power is switched ON.
b.3 Set Network Control to ON.
b.4 Select Network Config., and assure that Network Mode is set
to CPS/IDS. If necessary to change Network Mode from
Direct Net to CPS/IDS, select "Save ALL" before continuing.
Note: Monitor power-cycles automatically when "Save ALL"
is selected, if Network Mode is changed.
Caution
With the monitor in CPS/IDS mode, Network Config.
menu items report current settings of parameters in the
CPS or IDS with which the monitor is communicating.
Although the parameters cannot be initially programmed
into an unconfigured CPS or IDS, via the SC 7000 or SC
9000XL, they can be edited in the Network Config. menu.
Be extremely careful to avoid unintentionally changing
any item in the CPS/IDS configuration.
b.5 Go on to 9.
c If SC 7000 is to be connected directly to an I
monitor's ethernet port --
c.1 Assure that monitor is NOT mounted on docking station, and
set Network control to ON.
c.2 Select Network Config., and assure that Network Mode is set
to Direct Net. If necessary to change Network Mode from
CPS/IDS to Direct Net, select "Save ALL" before continuing.
Note: For first-time data entry, if parameters in Network
Config menu cannot be selected plug network cable into port
on right side-panel. Power-cycle monitor. Then disconnect
network cable and continue.
Monitor power-cycles automatically when "Save ALL" is
selected, if Network Mode is changed.
d Skip steps 9 through 15, and complete Table 4 in 3 before
continuing to the configuration procedure of 4.
9. Press Menu key to return to Main menu.
10. Select Monitor Setup → Unit Manager.
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
N
only through a
NFINTY
ETWORK
N
NFINTY
ETWORK
ASK-T898-03-7600
7k9kXLSM.ape.CD_ROM.fm/04-99/kaupp
via the

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sc 9000xl

Table of Contents