Siemens SC 7000 Service Manual page 133

Patient monitors
Table of Contents

Advertisement

SC 7000 and SC 9000XL Patient Monitors
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
800f5xxx
HEMOPOD / HEMOMED - related
800f6xxx
800fbxxx
NIBP - related
800fcxxx
80101xxx
Analog Out
80102xxx
80107xxx
RESP-related
80108xxx
8010dxxx
SPO2 - related
8010exxx
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
Description
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Cause
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting pneumatic
tubing.
If problem persists, replace
Main Processor Board
Subassembly.
Replace Main Processor
Board Subassembly.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting cable.
If problem persists, try
replacing sensor.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Service Manual
Action
117