Siemens SC 7000 Service Manual page 55

Patient monitors
Table of Contents

Advertisement

SC 7000 and SC 9000XL Patient Monitors
4.2.1 Removing Existing Label
4.2.2 Installing Language Label
4.2.3 Installing SIEMENS Metal
Logo Label
4.2.4 Installing Monitor Model
Label
4.3 Replacing Foot Pads
4.4 Removing / Installing
External Battery and
Support Components
4.4.1 External Battery (optional)
ASK-T898-03-7600
7k9kXLSM.c3.CD_ROM.fm/04-99/kaupp
To remove existing label, carefully peel up one corner of label and pull it
towards opposite corner.
Caution
Use extreme care to not damage underlying membrane keypad
when replacing Language Label. If membrane keypad is damaged
the Front Bezel/Lens Subassembly must be replaced.
1) Peel protective covering from section of new label under function keys.
2) Position label on membrane keypad so that it fits squarely in label
depression allowance on front bezel, and using a sweeping or rolling
motion to prevent air bubbles from becoming trapped under label,
press on label to secure it to membrane keypad.
3) Remove protective covering from remainder of language label, and
with a similar motion affix label along bottom of bezel.
Use this procedure if SIEMENS metal logo label not already installed on
front bezel.
1) Peel protective covering from new label.
2) Position in left rectangular depression in area near top left corner of
front bezel.
3) Press firmly, using a sweeping motion to prevent air bubbles from
becoming trapped under label.
1) Peel protective covering from appropriate new monitor model label.
Discard other new label.
2) Position NEW label in rectangular depression to right of SIEMENS
metal logo label in area near top left corner of front bezel.
3) Press firmly, using a sweeping motion to prevent air bubbles from
becoming trapped under label.
Two feet are on the bottom of the front bezel subassembly (see
3-10 on page
51, and four are on the bottom of the rear housing (see
Figure 3-2 on page
40). Pads are secured in foot wells by adhesive.
1) Remove all remnants of existing pad and adhesive from foot well.
2) Remove protective covering from adhesive surface of replacement pad.
3) Position replacement pad in foot well, and press firmly on pad to
secure it in well.
%
The external battery (
and 3-2) are contained in compartments on the rear of the monitor. It is
necessary to remove the back cover
The external battery compartment door is hinged to the compartment
&
housing on the right-hand end (facing the back of the monitor). The door
(
(
in
Figure
3-2) has been designed to lift the battery as an assist to the
ejection spring (
in
Figure
when the battery is being removed. It also secures the battery in the
compartment and assures good electrical connection to the battery
connector when closed.
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
in
Figure
3-1) and main battery (
$
only to replace the main battery.
3-2) at the opposite end of the compartment
Service Manual
!
%
in
Figure
in

in Figures
3-1
39