The following modules, PODs, and peripherals have no field-replaceable
parts or subassemblies.
• MultiMed and MultiMed 12 Pods, tcpO
• etCO
Module
2
• CPS/IDS (except for dc power cable in IDS-PS, and MIB Options in
IDS, which are field-replaceable)
• Hemo4 Pod / Hemo2 Pod, HemoMed Pod
• Remote Display
• PCMCIA Card
Remove all power sources, attached modules and cables, before
attempting to replace any items or open the monitor case.
&DXWLRQ
Open monitor only in a static-protected environment. Observe
standard precautions for protecting the equipment from static
electricity.
The following components and subassemblies can be replaced without
opening the monitor.
• rotary knob
• front bezel language and monitor model labels
• foot pads
• external battery (optional)
• external battery compartment door
• main battery
• power cable and speaker subassembly
• NP filters
• NP pump subassembly
• monitor handle
• rear housing side panels

The rotary knob (
in
Figure 3-10 on page
shaft of the optical encoder subassembly. It must be removed very carefully
if it is to be reinstalled.
To remove the knob, grip it very firmly with vise-grips or a similar tool, and
pull it straight out and off of the metal shaft. Avoid turning the knob.
Note: Placing a piece of cloth around the knob should prevent
scratching by the vise-grips, and allow the knob to be reused.
To install or reinstall a knob, align and firmly press knob onto shaft.
The Language Label kit ships with a set of language labels and SC 7000
and SC 9000XL model labels (see
SIEMENS metal logo label ships with the bezel (see
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
/CO
Pod, NeoMed Pod
2
2
!
and
battery ejector spring
51) is press fitted onto the metal


in
Figure 3-10 on page
in
7k9kXLSM.c3.CD_ROM.fm/04-99/kaupp
51). A
Figure
3-10).
ASK-T898-03-7600

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sc 9000xl

Table of Contents