Siemens SC 7000 Service Manual page 77

Patient monitors
Table of Contents

Advertisement

SC 7000 and SC 9000XL Patient Monitors
6.7 Record Key
6.8 Print Screen Key
6.9 NBP Start/Stop Key
6.10 Zoom Key
6.11 Help Key
6.12 Mark Key
7 ECG/RESP
Functions
7.1 ECG/RESP Test Setup
ASK-T898-03-7600
7k9kXLSM.c4.CD_ROM.fm/04-99/kaupp
1) Press the key and verify that a change occurs in the display.
2) Double-click key to return display to former status.
The Record key starts and stops a recording of limited duration when the
monitor is mounted on an IDS (or CPS).
1) With monitor mounted on IDS (or CPS) press Record key.
2) Verify that message "Recording Not Accepted" appears in message
field at bottom right-hand side of display.
Print Screen key initiates printing a copy of the display, on network printer.
1) Press Print Screen key twice.
2) Verify that message "Printer Unavailable, Waiting" appears in
message field at bottom right-hand side of display.
The NBP Start/Stop key initiates or terminates the inflation cycle for the
non-invasive blood pressure monitor function.
1) Press Menu key. Access Monitor Setup → Monitor Options → Speaker
Volumes, and set Attention Tone Volume to 50%.
2) Press NBP Start/Stop key.
3) Verify that monitor sounds a tone. (Cuff must not be plugged into cuff
connector.)
When a patient is being monitored, the Zoom key calls up a 1-hour trend
display for a quick overview of patient status.
1) Press Zoom key, and verify that "Fast Access Menu" displays.
2) Press Main Screen key to return.
1) Press Help key and observe that Main Help Menu appears on display.
2) Press Main Screen key to clear display.
The Mark key inserts parameter values with time and date stamp in the
tabular trends. Verify that Mark membrane switch is functional as follows:
1) Press Menu key and select Review → Trend Table.
2) Press Mark key and observe that time stamp on HR readout changes
from white to green and Mark symbol appears above stamp.
With the cable plugged into the monitor connect either a 3-lead, 5-lead, or
6-lead ECG cable from the Patient Simulator into the MultiMed POD.
1) Select HR parameter box and press rotary knob in to bring up ECG
menu.
• Set all ECG Lead settings at default values and remaining
parameters as follows:
--
ARR Monitoring
--
RESP Monitoring
--
Pacer Detection
• QRS Sync Marker
• Pulse Tone Source
• Pulse Tone Volume
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Basic
ON
ON
ON
ECG
10%
Service Manual
61