SC 7000 and SC 9000XL Patient Monitors
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
801b7001
ERROR: load_34010
801c7012
Test Failed! TAXI became Unavailable
801c9xxx
RECORDINGS - related
801caxxx
801cf014
801e1xxx
HEMOPOD / HEMOMED - related
801e2xxx
801e3004
801e7xxx
801e8xxx
801f2004
801f4004
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
Description
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Cause
34010 Failure
Replace Main Processor
Board Subassembly.
Expected condition,
Ignore.
not an error
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping Recorder.
If problem persists, replace
Main Processor Board
Subassembly.
Upgrade SW to ≥VC2.x.
Loaded box with
Rapid Pod Connects/
Disconnects
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Upgrade SW to ≥VC2.x.
Loaded box with
Rapid Pod Connects/
Disconnects
Try upgrading SW to
currently released version.
If error persists, replace
Main Processor Board
Subassembly.
Upgrade SW to ≥VC1.x.
Change to neonatal
category and TCGAS
reset
Upgrade SW to ≥VC1.x.
ST Zoom key reset
Service Manual
Action
121