SC 7000 and SC 9000XL Patient Monitors
Appendix E: Service Setup Instructions
1 Introduction
2 Initial Setup
2.1 Acccessing Service
Menu
ASK-T898-03-7600
7k9kXLSM.ape.CD_ROM.fm/04-99/kaupp
Following repair of an SC 7000 or SC 9000XL Patient Monitor, certain service
password-protected functions may need to be set for the specific customer
site, and for installed software versions ≥VC2 the monitor may also need to
be configured for use in DirectNet Mode.
Note: Turn the rotary knob to select a field or menu item, or
to change the default for a selected item. Press in on the
knob to access a menu or menu item, or to set a new default
for a selected item.
1. Firmly press ON/OFF key in lower left corner of front panel.
2. After the monitor has completed power-up sequence, as indicated by
illumination of "New Patient?" prompt, press rotary knob to select
"Yes" and bring up MAIN screen.
3. Select Monitor Setup → Monitor Options → Date & Time.
4. Set Time and Date as appropriate for customer site.
5. Press Menu key to return to Main menu.
6. Select Monitor Setup → Unit Manager.
7. Enter Clinical password (375), and then select Menu Setup on Unit
Manager menu.
8. Set Menu Time Limit to OFF.
9. Press Menu key to return to Main menu.
10. Go on to section 2.1.
1. Select Monitor Setup → Biomed → Service.
2. Enter Biomed password (4712).
3. Select Bedside Setup, and verify the following or set as required.
• Language setting is appropriate for customer site.
• Regulation is appropriate for customer site.
• Alarm Sounds set for either Default or Sirecust 900/1200 as preferred.
• Set Transport Brightness (display intensity when monitor is operating
on battery power) as requested by customer. (Lower intensity
conserves battery power.)
• Line Frequency is appropriate for customer site.
• Waveform Simulator is set OFF, for clinical use of Monitor.
4. Select
in upper left hand corner to return to previous menu.
5. Select Network Date and Time, and verify Current Date and Current
Time as appropriate for customer site. Set if necessary.
6. Select
to return to previous menu.
7. Set Printer Address, if necessary. Select Printer Address and turn
rotary knob to change number. Press knob in to enter new setting.
Note: "Printer Address" displays the fourth octet of the IP
address of the network printer (if any) to which printouts will
be sent from this monitor if in DirectNet Mode or from the
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Service Manual
139