Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
803a1xxx
12 LEAD POD - related
803a2xxx
803a3xxx
Serial Interface to Remote Keypad -
related
803a4xxx
8033b006
803e5xxx
SPO2 - related
803e6xxx
803e7xxx
803e8xxx
8035900b
IMRP Shutdown Timed Out!
80361xxx
12 LEAD POD - related
80362xxx
132
Description
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Cause
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting keypad.
If problem persists, swap
cable with known-good
keypad.
If problem persists, replace
Main Processor Board
Subassembly.
Upgrade SW to ≥VC2.x.
Remote View Display
Reset with Wedge
Try disconnecting/
reconnecting cable.
If problem persists, try
replacing sensor.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
ROM table on CPS or
Reinstall the software on
IDS is corrupt
CPS or IDS
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Action
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp