Siemens SC 7000 Service Manual page 72

Patient monitors
Table of Contents

Advertisement

Service Manual
56
4) Turn monitor bottomside up. See Figure 3-14.
(
5) Insert small screwdriver or similar tool (
15) into slot in tab (
6) Apply slight pressure in direction illustrated by heavy arrow in Figure 3-
15 to assure tab of rear housing is securely engaged in front bezel.
2
1
Figure 3-14 SC 7000 / SC 9000XL Patient Monitor Bottom View
(
7) Keeping pressure on lever (
security clip (
in Figure 3-15) vertically into space between bezel
and end of locking tabs in rectangular cutouts in bezel.
8) Rotate clip (hinge-like) so that offset tongues angle up under locking
tabs (to prevent tabs from being depressed) and surface of clip lies flat
on bezel.
9) Align peel-and-stick label (
tab slots (along line
to secure clip to bezel.
Note: Do NOT attempt to depress locking tabs for at least 24 hours.
+
Tape requires
24 hours to develop full adhesive strength.
1
Figure 3-15Installing Security Clip and Tape
10) Remove lever, and repeat steps 3 through 7 for other set of locking
tabs on bottom of bezel.
11) Reinstall left and right side panels (see Sections
ejector shaft cover (see section
12) Reconnect battery power and reinstall rear cover.
13) Functionally verify proper operation of the monitor before returning the
monitor to clinical service. Refer to
and Calibration"
.
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors

in Figure 3-14 and Figure 3-
in Figure 3-14) to use as a lever.

in Figure 3-15) , insert offset tongues of
!
"
in Figure 3-15) along edge of rectangular
in Figure 3-15), and press down firmly on tape
2
Section
5.1.2).
"Chapter 4: Functional Verification
7k9kXLSM.c3.CD_ROM.fm/04-99/kaupp
3
4
5.1.4
and 5.1.6), and
ASK-T898-03-7600