SC 7000 and SC 9000XL Patient Monitors
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
802abxxx
RECORDINGS - related
802acxxx
802adxxx
802aexxx
802b9xxx
RESP-related
802baxxx
802bbxxx
802bcxxx
802bf008
memory partition 2 is 91 percent used.
802d4002
shutdown not due to user. Could be low
battery.
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
Description
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Cause
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping Recorder.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly
Try upgrading SW to
currently released version.
If error persists, replace
Main Processor Board
Subassembly.
Memory utilization
Ignore
note
Monitor was shut
If runing monitor on internal
down without using
battery, ignore this
the user's front panel
message. Otherwise report
switch. Usually this
condition to customer
means that the
service
monitor was running
on internal batteries
and ran out of power.
Service Manual
Action
127