Service Manual
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
802dfxxx
SPO2 - related
802e0xxx
802e701c
Powering up the System...
802e701d
802ef011
trends completely cleared, new sofware
loaded
802f1xxx
Network Communications - related
802f2xxx
8030500d
b A a=B w=C d=D 0 e=E 0 s=F rx=G
ok=H flags=I df=J,K
80309xxx
8030axxx
128
Description
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Cause
Try disconnecting/
reconnecting cable.
If problem persists, try
replacing sensor.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Power was turned on. Ignore
Indicates that new
No action nessessary
software was
installed
Software warning
Ignore
Try disconnecting/
reconnecting cable(s)
between Monitor and
network wallbox.
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
SW error
Replace Main Processor
Board Subassembly.
Try upgrading SW to
currently released version.
If error persists, replace
Main Processor Board
Subassembly.
Action
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp