Siemens SC 7000 Service Manual page 131

Patient monitors
Table of Contents

Advertisement

SC 7000 and SC 9000XL Patient Monitors
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
800dbxxx
etCO2 Pod/Cartridge - related
800dcxxx
800ddxxx
HEMOPOD / HEMOMED - related
800dexxx
800dfxxx
800e0xxx
800e1xxx
NIBP - related
800e2xxx
800e4001
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
Description
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Cause
Try swapping out airway
adaptor with known-good
components.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Pod communications
Try disconnecting/
problem.
reconnecting cable.
Pod cable, Pod, or
If problem persists, swap
CPU board failed
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting pneumatic
tubing.
If problem persists, replace
Main Processor Board
Subassembly.
Upgrade SW to ≥VC2.x.
Remote View and
Wedge reset
Service Manual
Action
115