Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
80255xxx
80256xxx
8025bxxx
RS-232 Output - related
8025cxxx
80263xxx
80264xxx
80264001
Invalid status after RX_RESET!
rx_data=A
80264002
TX Empty should be set after
TX_RESET! tx_data=A
8026dxxx
Network Communications - related
8026exxx
124
Description
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Cause
Try upgrading SW to
currently released version.
If error persists, replace
Main Processor Board
Subassembly.
Ignore.
If problem persists, upgrade
SW to ≥VC2.x.
If problem persists, replace
Main Processor Board
Subassembly.
Recorder ASIC failed
Power cycle unit. If error
hardware test
persists, replace Main
Processor Board
Subassembly.
Recorder ASIC failed
Power cycle unit. If error
cold start test
persists, replace Main
Processor Board
Subassembly.
Try disconnecting/
reconnecting cable(s)
between Monitor and
network wallbox.
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
Action
ASK-T898-03-7600