SC 7000 and SC 9000XL Patient Monitors
4 Monitor Resets
5 Power On/Off Piezo Tone Fails to Sound.
Table 5-3 Power-off Alarm Malfunction
Symptom(s)
Piezo tone fails to
sound when monitor
powered ON or OFF,
or loses power.
6 Rotary Knob Malfunction.
Table 5-4 Rotary Knob Malfunction
Symptom(s)
Rotary knob fails to
properly select fields,
or pressing the knob in
fails to activate a menu
or select a default.
7 Fixed Key Fails to Function.
Table 5-5 Fixed Key Malfunction
Symptom(s)
A Fixed Key fails to
initiate change
ASK-T898-03-7600
7k9kXLSM.c5.CD_ROM.fm/04-99/kaupp
During startup and also during normal monitor operations, diagnostic
programs running in the background environment provide a constant
monitoring of critical hardware components and software functions.
In the event a critical component or software function is determined by the
software to be malfunctioning or there is no response to a requested task,
the diagnostics may force a full or partial main processor reset. Such events
can be identified by either a temporary loss of display or by a highly audible
watchdog alert tone. Monitor resets are typically recorded in the diagnostic
error log, which can be retrieved for interpretation and diagnosis by factory
trained experts.
Should a monitor exhibit symptoms of a reset, compare contents of the
diagnostic error log with the list of Error Messages and Diagnostic Codes
in Appendix C, for possible remedial action. Otherwise, record the
information on a copy of the
Chapter, and forward it along with the diagnostic error log to your respective
TSS center.
Possible Cause(s)
Main Processor PCB
malfunction
Possible Cause(s)
Optical Encoder
malfunction
Front Bezel PC Board
malfunction
Main Processor PCB
malfunction
Possible Cause(s)
Front Bezel Membrane
Switch or PC Board
malfunction
Main Processor PCB
malfunction
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
"Problem Report" on page 91
Troubleshooting and Remedial Action
1. Replace Main Processor PCB Subassembly.
2. If problem persists, contact TSS in Solna or Danvers.
Troubleshooting and Remedial Action
1. Replace Optical Encoder.
2. Replace Front Bezel PC Board.
3. If problem persists, replace Main Processor PCB
Subassembly.
4. If problem persists, contact TSS in Solna or Danvers.
Troubleshooting and Remedial Action
1. Replace Front Bezel/Lens Subassembly.
2. If problem persists, replace Front Bezel PC Board.
3. If problem persists, replace Main Processor PCB
Subassembly.
4. If problem persists, contact TSS in Solna or Danvers.