Siemens SC 7000 Service Manual page 118

Patient monitors
Table of Contents

Advertisement

Service Manual
4 Interface Plate
Figure B-6)Interface Plate Connectors (Rear View (Refer to Table B-4.)
Table B-4
Interface Plate Connectors Pinouts
Pin
RECORDER/ALARM (X7) EXT CRT (X5)
1
GND
2
+12VDC
3
Rec Tx
4
+12VDC
5
Diag Tx
6
+12VDC
7
Rec RTS
5 Recorder / Alarm Y Cable
Figure B-7Recorder / Alarm Y Cable (Accessory (Refer to Table B-5.))
Table B-5
Connector Pin No. Relay Input Wire Color SPDT Relay Output
1
2
3
4 - 8
9
10 - 14
102
;
Ext Red
VGND
Ext Grn
VGND
Ext Blu
VGND
GND
Recorder / Alarm Y Cable Connector Pinouts and Wire Color Code
Color Code (Alarm Cable)
TAN
NC
NC
NC
ORANGE
NC
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
;
Pin
RECORDER/ALARM (X7) EXT CRT (X5)
8
Rec CTS
9
Alarm Out
10
Rec GND
11
Rec GND
12
Rec Rx
13
Rec GND
14
Diag Rx
Circuit Status
Brown
RTN
Green
Inactive Open
White
Inactive Closed
H Sync
V Sync
Rem TxD
Rem RxD
Power Switch
Rem Audio
Rem Audio Ret
ASK-T898-03-7600
7k9kXLSM.apb.CD_ROM.fm/04-99/kaupp

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sc 9000xl

Table of Contents