Siemens SC 7000 Service Manual page 130

Patient monitors
Table of Contents

Advertisement

Service Manual
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
8000bxxx
8000cxxx
800c9009
atten OOR = A cal factor = B
800c9022
cal factor=A max_sample_value=B
800cd007
Loopback Active = A Expect Active =B
800d3xxx
Cardiac Output-related
800d4xxx
800d5xxx
800d6xxx
800d7xxx
Temperature - related (MultiMed Pod)
800d8xxx
800d9xxx
800daxxx
114
Description
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Cause
Software did not
Try disabling items
allocate enough
connected or disabling Auto
memory for user's
Dual View.
current setup.
If error persists, replace
Main Processor Board
Subassembly.
Speaker too quiet
Replace speaker
Speaker broken
Replace speaker
Hardware malfuction
Check "Nurse Call" circuit.
in "nurse call" circuit.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Pod communications
Try disconnecting/
problem.
reconnecting cable.
Pod cable, Pod, or
If problem persists, swap
CPU board failed
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Action
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp